VNA Automatic Calibration via SCPI
================================================

Introduction
============

This chapter describes how to automate the calibration process of a Vector Network Analyzer (VNA) 
using **Standard Commands for Programmable Instruments (SCPI)** and Python. 

Calibration is the most critical step in RF measurements. It mathematically removes the 
systematic errors (Directivity, Source Match, and Reflection Tracking) introduced by the VNA 
hardware and cables, shifting the measurement plane to the end of the test port.

Objectives
----------

This tutorial serves two primary purposes:

1. **Practical Application**: To perform a "Full One-Port" (FOPort) automatic calibration 
   required for accurate reflection measurements (S11) up to 18 GHz.
2. **Skill Development**: To provide a hands-on example of how to communicate with high-end 
   test equipment using the PyVISA library and raw SCPI strings.

System Requirements
===================

* **Hardware**: Rohde & Schwarz ZNL series VNA and a compatible Automatic Calibration Unit (e.g., ZN-Z151).
* **Connection**: Ethernet (TCP/IP) via HiSLIP protocol.
* **Libraries**: ``pyvisa``, ``numpy``.

Implementation
==============

The following Python script initializes the instrument, sets the frequency sweep from 100 MHz 
to 18 GHz, and triggers the internal AutoCal routine.

.. code-block:: python
   :linenos:

   import pyvisa

   # Initialize communication
   rm = pyvisa.ResourceManager()
   vna = rm.open_resource('TCPIP0::10.0.0.5::hislip0::INSTR')
   vna.timeout = 120000  # Calibration hardware takes time to switch internal states

   # Reset the instrument to default state
   vna.write("*RST")

   # Set frequency range: 100 MHz to 18 GHz
   vna.write("SENSe1:FREQuency:STARt 100e6")
   vna.write("SENSe1:FREQuency:STOP 18e9")
   vna.write("SENSe1:SWEep:POINts 1601")

   # Run autocal on port 1 using factory characterization ('')
   # The VNA automatically detects the CalUnit connected to Port 1
   vna.write("SENSe1:CORRection:COLLect:AUTO '', 1")
   
   # *OPC? is vital: it pauses the script until the VNA hardware is finished
   vna.query("*OPC?") 

   # Verify calibration status
   state = vna.query("SENSe1:CORRection:STATe?").strip()
   print(f"Calibration active: {state}")  # 1 = Success/Active

   # Save the resulting calibration data to the internal storage
   vna.write("MMEMory:STORe:CORRection 1, 'AutoCal_Port1.cal'")

   vna.close()
   rm.close()

   # Later load with
   #vna.write("MMEMory:LOAD:CORRection 1, 'AutoCal_Port1.cal'")

Command Breakdown
=================

* ``SENSe1:CORRection:COLLect:AUTO '', 1``: The empty string tells the VNA to use the factory 
  data stored inside the CalUnit. The ``1`` specifies the physical Port 1 of the VNA.
* ``*OPC?``: Stands for "Operation Complete." It prevents the script from closing the 
  session before the physical switches inside the CalUnit have finished cycling.






Extracting and Verifying Calibration Error Terms
================================================

Once a calibration is performed and saved, it is often necessary to verify the quality of the 
calibration or extract the error terms for offline analysis. The following script demonstrates 
how to load a previously saved calibration file and retrieve the **Directivity**, **Source Match**, 
and **Reflection Tracking** coefficients.

.. code-block:: python
   :linenos:

   import pyvisa
   import numpy as np
   import skrf as rf

   rm = pyvisa.ResourceManager()
   vna = rm.open_resource('TCPIP0::10.0.0.5::hislip0::INSTR')
   vna.timeout = 100000

   # Reset and load the specific calibration file
   vna.write("*RST")
   vna.write("MMEMory:LOAD:CORRection 1, 'AutoCal_Port1.cal'")

   # Verification Queries
   print(f"Is AutoCal: {vna.query('SENSe1:CORRection:DATA:PARameter? ACAL')}")
   print(f"Cal Date: {vna.query('SENSe1:CORRection:DATE?')}")

   # Set to HOLD mode to ensure data consistency during readout
   vna.write("SENSe1:SWEep:MODE HOLD")

   # Request raw error term data
   D_raw = vna.query("SENSe1:CORRection:CDATa? 'DIRECTIVITY', 1, 0")
   S_raw = vna.query("SENSe1:CORRection:CDATa? 'SRCMATCH', 1, 0")
   R_raw = vna.query("SENSe1:CORRection:CDATa? 'REFLTRACK', 1, 0")

   # Data Transformation: From String to Complex Array
   D_complex = np.array(D_raw.split(','), dtype=float)
   D_complex = D_complex[0::2] + 1j * D_complex[1::2]

   S_complex = np.array(S_raw.split(','), dtype=float)
   S_complex = S_complex[0::2] + 1j * S_complex[1::2]

   R_complex = np.array(R_raw.split(','), dtype=float)
   R_complex = R_complex[0::2] + 1j * R_complex[1::2]

   print(f"Points Extracted: {len(D_complex)}")
   print(f"First Directivity Point: {D_complex[0]}")

   vna.write("SENSe1:SWEep:MODE CONTinuous")

   #Converting error terms to Network objects for visualizing them using Sci kit RF

   freq = rf.Frequency(start=100e6, stop=18e9, npoints=1601, unit='Hz')
   ntwk_D = rf.Network(frequency=freq, s=D_complex, name='Directivity')
   ntwk_S = rf.Network(frequency=freq, s=S_complex, name='Source Match')
   ntwk_R = rf.Network(frequency=freq, s=R_complex, name='Refl Tracking')
   
   
   plt.figure(figsize=(15, 5))
   
   plt.subplot(1, 3, 1)
   ntwk_D.plot_s_db()
   plt.title('Directivity (ED)')
   
   plt.subplot(1, 3, 2)
   ntwk_S.plot_s_db()
   plt.title('Source Match (ES)')
   
   plt.subplot(1, 3, 3)
   ntwk_R.plot_s_smith()
   plt.title('Reflection Tracking (ER)\n(Smith Chart)')
   
   plt.tight_layout()
   plt.show()


   vna.close()
   rm.close()

Understanding the Data Transition: D_raw to D_complex
====================================================

The transition from the raw string returned by the VNA to a usable complex mathematical array 
in Python happens in three distinct stages:

1. String Splitting and Type Casting
------------------------------------
When the VNA executes ``CDATa?``, it returns a single, massive string of comma-separated values 
(e.g., ``"0.012,-0.005,0.011,..."``). 

* ``D_raw.split(',')``: Breaks the string into a Python list of individual substrings.
* ``np.array(..., dtype=float)``: Converts those substrings into 64-bit floating-point numbers.


2. De-interleaving with Slicing
-------------------------------
VNA hardware communicates complex data in an **interleaved** format: 
``[Real_1, Imag_1, Real_2, Imag_2, ...]``. To perform vector math, we must separate these.

* ``D_complex[0::2]``: This uses NumPy slicing to start at index 0 and take every **2nd** element. This captures all **Real** components.
* ``D_complex[1::2]``: This starts at index 1 and takes every **2nd** element. This 
  captures all **Imaginary** components.

3. Complex Vector Reconstruction
--------------------------------
The final step uses the imaginary unit ``1j`` (the Python equivalent of :math:`i`) to 
rebuild the vectors:

.. math::
   D_{complex} = Real + j \cdot Imaginary

By adding the two sliced arrays together with the ``1j`` multiplier, NumPy creates a 
single array of **complex128** objects. This allows you to calculate Magnitude 
(``np.abs()``) and Phase (``np.angle()``) directly, which is essential for analyzing 
systematic errors at high frequencies like 18 GHz.



Visualization of Error Terms
============================

After extracting the complex coefficients, we can visualize the VNA's internal state. These plots provide insight into the quality of the calibration and the physical characteristics of the test setup.

.. figure:: media/Autocal.png
   :align: center
   :alt: VNA Calibration Error Terms
   :width: 800px

   **Figure 1**: Log-Magnitude plots of Directivity and Source Match, alongside a Smith Chart representation of Reflection Tracking.

Data Interpretation
-------------------

* **Directivity (ED)**: Represents the leakage within the VNA. A value moving from -50 dB toward -15 dB at 18 GHz is typical, showing reduced isolation at higher frequencies.
* **Source Match (ES)**: Shows the reflections looking back into the VNA port. The values between -35 dB and -18 dB indicate a healthy match for a standard cable assembly.
* **Reflection Tracking (ER)**: The "spiral" seen on the Smith Chart is the result of the phase delay introduced by the cables and PCB traces. At 18 GHz, the signal undergoes multiple full rotations of phase, which the VNA mathematically tracks to shift the measurement plane to your device.