MRI Hardware#
Overview#
Magnet types: superconducting, permanent, resistive; cryostat & cryogens
Gradient coils, gradient amplifiers, slew rate, active shielding
RF chain: body transmit, surface/array receive coils, pre-amps, T/R switches
Shim coils (passive & active)
Console electronics: waveform generators, ADC/DAC, clocks
Patient table, positioning, physiological gating hardware
RF shielding, magnetic shielding, cryogenic venting, Faraday cage
Safety subsystems: quench pipe, interlocks, SAR/DB1 monitors
Note
1 liter of liquid helium costs as much as good whiskey (analogy) — around 50 pounds.
Kuprov
Main magnet and shielding#
Shim systems#
Even 0.01% deviation in current can lead to 20 ppm inhomogeneity in 7T.
Subject inhomogeneity (from humans):
small (< 2 ppm)
subject-dependent
mostly static
dynamic (breathing and motion)
\(\lambda\): magnetic susceptibility
\(\lambda < 0\): diamagnetic (water, copper)
\(\lambda > 0\): paramagnetic (air, niobium)
\(\lambda \gg 0\): ferromagnetic (iron, nickel)
Magnetic field in each of those materials, taken separately, would still be homogeneous; however, inhomogeneities arise where two materials with different susceptibilities meet. Air–water interfaces give the biggest disturbances in magnetic field, which is typical for humans (nose and ear canals with air, other parts with water).
Shimming is the process of adding magnetic fields to cancel inhomogeneities.
Mathematically, the inhomogeneous field can be approximated by the homogeneous field we want plus a sum of inhomogeneity terms described by basis functions (commonly spherical harmonic functions). The more terms available, the better the approximation; in practice, most scanners implement first- and second-order harmonics.
Nice property: spherical harmonic fields can be generated inside the magnet using specifically designed coils.
First-order fields are generated by X, Y, Z gradient coils.
Second-order fields are generated by at least five additional coils.
Spherical harmonic shim coils are typically superconducting.
Gradient coils#
Primary function#
Encode spatial information: - Slice selection - Phase encoding - Frequency encoding
Secondary functions#
Spin de-phasing to minimize artifacts
Spin re-phasing after slice selection
Spin pre-phasing before readout
Sensitize/de-sensitize images to motion
Gradients are a special kind of inhomogeneous magnetic field where the z-component varies linearly along a specific direction. For all X, Y, Z gradients, the magnetic field component produced by the gradient points strictly along \(\hat{k}\) (the \(z\)-axis). Gradients are not fields perpendicular to \(B_0\); otherwise, the net field would be a vector sum leading to an oblique gradient.
Where:
\(B\): magnetic field from the gradient
\(G_z\): points along \(z\)
\((x)\): varies with \(x\)
\(G_x\): X-gradient amplitude
\(x\): position relative to isocenter
More generally, with position vector \(\vec{r}\):
Larmor frequency of a nucleus at position \(\vec{r}\) is:
Because \(B_0\) is orders of magnitude stronger than gradient fields, the Larmor frequency is governed mainly by \(B_z(\vec{r})\), with perpendicular components negligible:
Ideal gradient coil requirements (constant spatial derivative):
These can be met only within a finite region, the linearity volume (LV), often specified as the diameter of a spherical volume (DSV) with linearity radius \(r_{LV}\). A key design goal is to maximize DSV.
Another challenge is eddy currents induced in conductive structures by time-varying fields (Faraday’s law). A common solution is a secondary shielding coil placed outside the main gradient coil (actively shielded gradients).
Design methods:
Gradient descent
Least-squares methods
Simulated annealing
Z gradient#
Maxwell pair coil
Linear B-field variation along \(z\)
Highly uniform in the \(xy\) plane
Currents are equal and opposite
XY gradients#
Saddle/Golay coil pairs
Linear B-field variation along \(x\) or \(y\)
Highly uniform in the \(yz\) or \(xz\) plane
Innermost turns produce the gradient field
RF chain#
RF coils are the lenses of the MRI system.
RF must produce a time-varying excitation \(B_1(t)\) with:
Components that rotate near the resonant frequency \(\omega_0\)
Components perpendicular to the static field \(B_0\)
Progression of coil types:
Single loop (simplest) → Helmholtz (2-coil) → \(N\)-coil (solenoid)
These are linearly polarized (LP): \(B_1\) oscillates in one dimension, so only half the energy couples to the spins.
They are replaced in many transmit applications by quadrature / CP (circularly polarized) coils (two orthogonal fields). LP coils remain widely used as receive-only coils; quadrature detection improves SNR by a factor of \(\sqrt{2}\).
General RF coil design goals:
Maximize transmit \(B_1^+\) within the sample per unit input current
Minimize electric field \(E\) within the sample per unit input current
For volume coils, achieve homogeneous \(B_1^+\) over as large a volume as possible
Flip angle:
For a given flip angle, increasing \(B_1^+\) shortens the pulse duration \(\tau\), which helps minimize off‑resonance artifacts.
Sample power absorption:
Where \(\sigma\) is conductivity and \(v\) is sample volume.
Electronics#